Drager Oxylog Ventilator

Drager Oxylog Ventilator

Product Details:

X

Product Description

Specifications :

MODEL

Oxylog 2000

MONITORED PARAMETERS
Pressure
Pip (MONITORED PARAMETERS)Yes
Map (MONITORED PARAMETERS)Yes
Peep (MONITORED PARAMETERS)Yes
Volume
Tidal (Volume)Yes
Minute (Volume)Yes
Respiratory rate (Volume)Yes
Inspiration time (MONITORED PARAMETERS)Yes
Additional (Alarms, high/low)CPAP pressure level, Vt expiratory
INTERNAL BATTERYYes
Type (number) (ALERTS)Ni-Cd or alkaline manganese (not specified)
Operating time, hr (Spontaneous/cpap)Up to 6 (Ni-Cd), up to 4 (alkaline manganese)
Rechargeable (POWER NEEDED)Yes
Recharging time, hr (POWER NEEDED)8 maximum
DIMENSIONS (HXWXD) CM,(IN) (ALERTS)12.3 x 21.5 x 20.8 (4.5 x 8.5 x 8.2)
WEIGHT, kg (lb) (DISPLAY)4.3 (8.5)
Service
Factory/on-site (Service)Yes/yes
3rd-party service (Service)Yes
Parts availability (ALERTS)Yes
Number sold (PURCHASE INFORMATION)
REGULATORS
Tidal volume, ml (CONTROLS)100-1,500
Inspiratory flow, l/min (REGULATORS)1-4
Inspiratory pressure, cm h2o (REGULATORS)20-60
Respiratory rate, breaths/min (REGULATORS)1-5
Inspiratory time, sec (CONTROLS)Depends on set frequency and I:E ratio
Expiratory time, sec (REGULATORS)Depends on set frequency and I:E ratio
Ie ratio (REGULATORS)1:3 to 2:1
Fio2, % (CONTROLS)60 or 100
Peep/cpap, cm h2o (REGULATORS)0-15
Trigger system (REGULATORS)Flow
100% o2 (REGULATORS)Yes
Additional (Alarms, high/low)CPR mode: inspiration phase not interrupted; more O2 exchange
POWER NEEDED, VAC (ENERGY)100-240
Current, amps (REGULATORS)0.7, maximum 1.6
POWER CONSUMPTION, W (OTHER AVAILABLE ROTORS)2.8 for ventilation, 3.6 for battery charging
PNEUMATIC POWER (Volume)
CompressorNot specified/not integrated
Compressed gases (EQUIPMENT ALARMS)Medical grade air or O2
Pressure ranges (EQUIPMENT ALARMS)2.7-6 bar, 39-90 psi
PATIENT ALARMS
Low minute volume (REGULATORS)Yes
Low inspiratory pressure (PATIENT ALARMS)Yes
Over pressure (MONITORED PARAMETERS)Yes
Apnea (CONFIGURATION)Yes
High continuous pressure/occlusion (EQUIPMENT ALARMS)Yes
High respiratory rate (PATIENT ALARMS)Yes
High minute volume (PATIENT ALARMS)Yes
Breathing circuit disconnect (PATIENT ALARMS)Yes
ALERTS (CONFIGURATION)
Gas supply malfunction (ALERTS)Yes
Power loss (ALERTS)Yes
Vent inoperative (EQUIPMENT ALARMS)Yes
Low battery (EQUIPMENT ALARMS)Yes
SELF-TESTYes
INTERFACING
Report generation
DISPLAYS (STAINING POSTS)4 x 20 LCD, 9 languages; adjustable
DATA DISPLAYED (REGULATORS)Settings, alarms, measured values, pressure and flow values and alarms
MODES OF OPERATION
Assist/control
Volume breaths (Assist/control)Yes
Simv
Volume breaths (Assist/control)Yes
Pressure breaths (Assist/control)Pressure limited
Spontaneous/cpap
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
WHERE MARKETEDWorldwide

 

Enter Buying Requirement Details
Email Id
Mobile number


Back to top