Drager Babylog 8000 Ventilator

Drager Babylog 8000 Ventilator

Product Details:

X

Product Description

Technical Specifications :

  • Conventional ventilation: continuous flow, pressure-limited, time-cycled
  • Triggered ventilation: SIMV, SIPPV, PSV1), leak adapted
    - Trigger: Flow/volume trigger, leak adapted
    - Trigger delay: approx. 40-60 ms
  • High Frequency Ventilation: CPAP+HFV, IMV+HFV
    - Frequency: 5 to 20 Hz
  • Volume Guarantee Ventilation: SIMV+VG, SIPPV+VG, PSV+VG
  • Oxygen mixer loss (bleed flow): 0 (zero) L/min

Settings

  • Inspiratory oxygen concentration: 21-100 vol.% oxygen
  • Peak inspiratory pressure: 10 to 80 mbar
  • PEEP/CPAP: 0 to 25 mbar
  • Maximal frequency: 200 bpm
  • Inspiratory time: 0.1 to 2 seconds
  • Expiratory time: 0.2 to 30 seconds
  • Inspiratory flow: 1 to 30 L/min
  • Base Flow (VIVE): 1 to 30 L/min

Monitoring

  • Flow monitoring: at the Y-piece, integrated
  • Volume monitoring: at the Y-piece, integrated
  • Lung function monitoring: compliance, resistance C20/C
  • Method: linear regression analysis
  • FiO2 monitoring: integrated
  • Real-time curves: Flow and pressure, integrated
  • Inspiratory oxygen concentration: 21-100 vol.% oxygen
  • Peak pressure: to 99 mbar
  • Mean airway pressure: to 99 mbar
  • Graphic trends: 6 parameters, integrated
  • Logbook: record of up to 100 alarms

 

Enter Buying Requirement Details
Email Id
Mobile number


Back to top